Sterilization

| 0

Please consider our TechNote of the sterilization methods of the mp6 micropumps. For ebeam and gamma sterilization we had a customer project where a system with the pump was tested before and after the sterilization processes. No impact of the … Weiter

Humidity

| 0

The mp6 micropumps have been tested running at humidities from 95% to ~10%. These shortterm tests can not give hints regarding long-term usage, as water can appear at electric connections inside Molex connector and beneath the lid where the actuators … Weiter

Operating gas with particles

| 0

We don’t have data on particle effects with the mp6-gas micropump. The mp6-gas is also possible to pump liquids. In general, the mp6-gas is just a standard mp6 pump, only tested especially for air performance, not every mp6 achieves the … Weiter

Leaking of the tubing

| 0

It is possible to push a host clamp over the attached tubing end to clamp it tightly to the connector. For a good sealing we recommend our mp-hc which gives a good sealing regarding leaking.  Another option is to use … Weiter

Electronic connection of the micropump

| 0

The pump cable has to be inserted with the contact surfaces orientated down into the plug. Both components are firmly connected by closing the white plug. When disconnecting the pump, the board connector must first be opened! Verify yourself that … Weiter

Value pair of voltage and frequency

| 0

The optimum frequency should be selected based on the medium: 100 Hz for water, 300Hz or higher for gases, 50Hz or lower for oils. Thereafter, the position of the working line of the pump (straight line from maximum pressure to … Weiter

Pumping blood?

| 0

In principle it is possible to pump blood with the mp6 micropump. There might be some damaging of blood cells (red blood particles), but only with certain pump settings. The optimal pump settings has to be tested. Our recommendation is … Weiter