Material in contact with fluids

| 0

All wetted materials inside the standard mp6 are PPSU.. Please note that the mp6 is made with material of two grades of PPSU; natural-transparent and black. The lid is black. Radel R-5000 NT is the natural-transparent material and Radel R-5100 … Weiter

Test equipment

| 0

We are providing our testing and measurement equipment data setups on request, allowing to test and evaluate the micropumps in the same way as at our product release test. When selecting sensors, please be aware, whatever sensor you apply, make … Weiter

Sterilization

| 0

Please consider our TechNote of the sterilization methods of the mp6 micropumps. For ebeam and gamma sterilization we had a customer project where a system with the pump was tested before and after the sterilization processes. No impact of the … Weiter

Humidity

| 0

The mp6 micropumps have been tested running at humidities from 95% to ~10%. These shortterm tests can not give hints regarding long-term usage, as water can appear at electric connections inside Molex connector and beneath the lid where the actuators … Weiter

Lifetime of the micropump mp6

| 0

We guarantee a lifetime of minimum 5.000 continuous working hours under lab conditions for the micropump mp6 series (>10.000 hours have been achieved without any notable failures). These values have been achieved by using a micropump from the mp6 series … Weiter